HomeGas detectors and meters4-gas calibration mixture 60DAQUAD-004 — 60 l cylinder
← Back to Gas detectors and meters
4-gas calibration mixture 60DAQUAD-004 — 60 l cylinder

4-gas calibration mixture 60DAQUAD-004 — 60 l gas cylinder

Certified calibration gas mixture for calibration, adjustment and bump testing of four-gas detectors. One cylinder serves all four measuring channels: hydrogen sulphide, carbon monoxide, methane and oxygen.

Mixture composition

ComponentConcentration
Hydrogen sulphide (H₂S)25 ppm
Carbon monoxide (CO)100 ppm
Methane (CH₄)2.2% vol.
Oxygen (O₂)18% vol.
Nitrogen (N₂)balance

This is the composition most often used for four-gas instruments — the reduced oxygen level lets you verify the O₂ deficiency alarm, while methane and the toxic gases check the remaining channels in a single pass.

What it is for

  • Bump test before use — a short check that the sensors and alarms respond.
  • Adjustment — correcting the instrument readings to the reference values.
  • Calibration — verifying the readings after adjustment.
  • Works with docking and calibration stations, including Honeywell MicroDock II and IntelliDoX.

Technical data

ParameterValue
Product code60DAQUAD-004
Gas volume60 l
CylinderLuxfer aluminium, 1.7 l water capacity
Cylinder weight0.76 kg
Filling pressure500 psi (approx. 34 bar)
Test pressure105 bar
Cylinder standardISO 7866
Transport classificationUN1956, class 2.2 — compressed gas
Shelf life2 years from the date of manufacture
Valvestandard, for calibration regulators
Cylinder markingscylinder number, date of manufacture and expiry date

Safety and storage

  • Use only with a compatible pressure regulator. The valve, regulator and fittings must be clean.
  • Open the valve slowly. Work in a well-ventilated room.
  • Store upright and secured against falling over.
  • Markings: H280 — contains gas under pressure, may explode if heated. P403 — store in a well-ventilated place.
  • Dispose of in line with local regulations. The cylinder is intended for recycling.

Every cylinder carries its own serial number, mixture code, date of manufacture and expiry date on the label — this lets you tie the reference unambiguously to the calibration record. The transport classification and hazard markings are printed directly on the cylinder.

Choosing the mixture

The composition given here suits typical four-gas detectors. If your instruments measure other gases or work to different alarm thresholds, we will prepare a mixture for that specific configuration — tell us which sensors your instruments carry.

Not sure how often to calibrate? We explain it on the gas detector servicing page — we service equipment at your site or in our laboratory in Suchy Dwór near Gdynia.